Datex-Ohmeda, Inc. recalls
9 recalls in our records from 2018 to 2026: 9 FDA medical devices.
Recalls with full detail pages
-
Datex-Ohmeda Inc medical device recall: Giraffe Stand-alone Infant Resuscitation System
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended…
Earlier and other recalls
These link to the official agency notice.
-
Datex-Ohmeda, Inc. medical device recall: Aisys Official notice ↗
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system…
-
Datex--ohmeda, Inc. medical device recall: GE Healthcare Giraffe Omnibed Carestation CS1 Official notice ↗
Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical…
-
Datex-Ohmeda, Inc. medical device recall: Aespire 7100 Official notice ↗
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
-
Datex--ohmeda, Inc. medical device recall: GE Heathcare Giraffe OmniBed Official notice ↗
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters…
-
Datex-Ohmeda, Inc. medical device recall: EVair air compressor Official notice ↗
GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m…
-
Datex--ohmeda, Inc. medical device recall: Giraffe OmniBed Carestation CS1 Official notice ↗
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside…
-
Datex-Ohmeda, Inc. medical device recall: The Neonatal Flow Sensor Cable used on devices CARESCAPE R860 Official notice ↗
The Neonatal Flow Sensor Cable may be missing the connector housing exposing the wires in the cable.
-
Datex-Ohmeda, Inc. medical device recall: CARESCAPE R860 ventilators with software version 10SP05 Product… Official notice ↗
Potential for the display processor to experience an unexpected failure.