Intersurgical Inc recalls
8 recalls in our records from 2017 to 2026: 8 FDA medical devices.
Recalls with full detail pages
-
Intersurgical Inc medical device recall: i-view video laryngoscope. Model Number: 8008000. The i-view video…
Potential for faulty devices as a result of depleted batteries.
Earlier and other recalls
These link to the official agency notice.
-
Intersurgical Inc medical device recall: One-Piece Guedel Airway Official notice ↗
Devices could potentially be occluded or partially occluded.
-
Intersurgical Inc medical device recall: One-piece Guedel airway Official notice ↗
Potential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue irritation,…
-
Intersurgical Inc medical device recall: Intersurgical Solus Standard Official notice ↗
Printing error on the device, size 3 Solus Standard laryngeal mask airway has been incorrectly marked with size 4 information (includes the number 4 as a size…
-
Intersurgical Inc medical device recall: Smoothbore breathing system with exhalation port 22 mm Device REF #:… Official notice ↗
The device was assembled with the incorrect exhalation port which can cause gas leakage.
-
Intersurgical Inc medical device recall: Double Swivel Elbow with Bronchoscopy Port 15MM OD Official notice ↗
The port may crack in the swivel elbow where the flip cap is inserted, and these cracks could possibly leak if they reach a significant size.
-
Intersurgical Inc medical device recall: Flexible Smoothbore Breathing Circuit with Exhalation Port 22mm Official notice ↗
When assembling an HME or Filter on this breathing circuit the tubing may be pushed up and inadvertently cover or block the exhalation port. This could potentially pose…
-
Intersurgical Inc medical device recall: Solus Flexible wire-reinforced laryngeal mask airways under the… Official notice ↗
A defect in the wire-reinforced tubing may result in a partial of total occlusion when the cuff is over-inflated.