RecallRadar
FDA · Medical devicesClass I

Mercury Enterprises, Inc. dba Mercury Medical medical device recall: Neo-Tee T-Piece Resuscitator

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
95967
Recall initiated
Dec 10, 2024
Classified by FDA
Jan 10, 2025
Posted in Enforcement Report
Jan 22, 2025
Recalling firm
Mercury Enterprises, Inc. dba Mercury Medical
Firm location
Clearwater, FL, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Dec 10, 2024. FDA classified it 31 days later, on Jan 10, 2025. Distribution reported by FDA: Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands Mercury Enterprises, Inc. dba Mercury Medical has 3 recalls in our records.

Recalled products (1)

  • Z-0821-2025 Class I Ongoing

    Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842

    Codes / lots: Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805; 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808; 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809; 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810; 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811; 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814; 1050832/10641043508329, 30641043508323/2429050832, 242985083; 1050839/10641043508398, 30641043508392/2430350839; 1050840/10641043508404, 30641043508408/2429050840,…

    Quantity: 37370

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 95967
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.