Micro Therapeutics, Inc. medical device recall: Medtronic Pipeline Vantage Embolization Device with Shield Technology
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 96168
- Recall initiated
- Jan 30, 2025
- Classified by FDA
- Mar 5, 2025
- Posted in Enforcement Report
- Mar 12, 2025
- Recalling firm
- Micro Therapeutics, Inc.
- Firm location
- Irvine, CA, United States
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Letter
- Products in this event
- 2
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Jan 30, 2025. FDA classified it 34 days later, on Mar 5, 2025. This recall event covers 2 separate products, each with its own FDA recall number. Distribution reported by FDA: Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile,… Micro Therapeutics, Inc. has 15 recalls in our records.
Recalled products (2)
Z-1245-2025 Class I Ongoing
Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30, PED3-027-450-12, PED3-027-450-14, PED3-027-450-16, PED3-027-450-18, PED3-027-450-20, PED3-027-450-25, PED3-027-450-30, PED3-027-450-40, PED3-027-500-14, PED3-027-500-16, PED3-027-500-18, PED3-027-500-20, PED3-027-500-25, PED3-027-500-30, PED3-027-500-40, PED3-027-550-16, PED3-027-550-18, PED3-027-550-20, PED3-027-550-30, PED3-027-550-40, PED3-027-550-50, PED3-027-600-16, PED3-027-600-18, PED3-027-600-20, PED3-027-600-30, PED3-027-600-40, PED3-027-600-50
Codes / lots: All unused devices. Product lots including and between B317266 and B817292. REF/UDI-DI/Lot: PED3-027-350-12/00847536035263, 00847536030770/B409106, B637287, B700548; PED3-027-350-14/00847536035287, 00847536030787/B409234, B700550, B794880; PED3-027-350-16/00847536035294, 00847536030794/B409692, B697962, B700551, B795136, B795142, B809808; PED3-027-350-20/00847536035317, 00847536030817/B410645, B558444, B623124, B672921, B700552, B777630; PED3-027-350-25/00847536035324, 00847536030824/B410718, B672530, B700553, B771934, B771938; PED3-027-400-12/00847536030862, 00847536035362, 00847536036680/B321585, B336094, B349668,B371137, B409814, B414591, B437740, B482271, B594119, B597085, B598465, B612514, B616043, B616052, B647374, B649853, B655648, B658620, B658685, B666385, B666418, B671805, B720634, B720635, B744619, B744621, B756639, B765446, B765448, B780611, B786108;…
Quantity: 18,245
Z-1246-2025 Class I Ongoing
Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2
Codes / lots: Instructions for Use, Version: Rev. A (01/2024). REF/UDI-DI/Lot: PED3-021-350-16/00847536030701, 00847536035195, 00847536036529/B317460, B327449, B335184, B340948, B349032, B389162, B392918, B424454, B430015, B495821, B497938, B547396, B550307, B551615, B559033, B603377, B623564, B656863, B656887, B658904, B692346, B715057, B715058, B715541, B726894, B735821, B735823, B756029, B756030, B760210, B760211, B768202, B768387, B769453, B770633, B807494; PED3-021-250-14/00847536030411, 00847536034907, 00847536036239/B317936 B332767, B333492, B333536, B334545, B376967, B388041, B391185, B399354, B416713, B431395, B452416, B457963, B467691, B476239, B564173, B564970, B574472, B580923, B629372, B630998, B670305, B673035, B699576, B707568, B709193, B736698, B736893, B739090, B739091, B795110, B796586, B800827, B801770; PED3-021-350-20/00847536030725, 00847536035218, 00847536036543/B319436,…
Quantity: 7820
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
- Official notice
- FDA Enforcement Report, event 96168
- Raw record
- openFDA API (JSON)
- Dataset last updated by FDA
- Nov 3, 2026
- Imported by RecallRadar
- Nov 4, 2026
Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.