RecallRadar
FDA · Medical devicesClass I

Thoratec LLC medical device recall: HeartMate Mobile Power Unit: REF: 107754

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
96471
Recall initiated
Mar 13, 2025
Classified by FDA
Apr 17, 2025
Posted in Enforcement Report
Apr 23, 2025
Recalling firm
Thoratec LLC
Firm location
Pleasanton, CA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Mar 13, 2025. FDA classified it 35 days later, on Apr 17, 2025. Distribution reported by FDA: Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA, KS, AL, CO, ME and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CZECH REPUBLIC, FINLAND, FRANCE, GEORGIA, GERMANY, HONG KONG, ISRAEL, ITALY, JAPAN, KUWAIT, LEBANON, LITHUANIA,… Thoratec LLC has 17 recalls in our records.

Recalled products (1)

  • Z-1570-2025 Class I Ongoing

    HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.

    Codes / lots: UDI-DI: 05415067038234. MPUs distributed between April 2024 and February 2025. Serial Numbers: 20226263, 20226264, 20226265, 20226266, 20226267, 20226268, 20226269, 20226270, 20226271, 20226272, 20226273, 20226274, 20226275, 20226276, 20226277, 20226278, 20226279, 20226280, 20226281, 20226282, 20226283, 20226284, 20226285, 20226286, 20226287, 20226288, 20226289, 20226290, 20226291, 20226292, 20240622, 20240623, 20240624, 20240625, 20240626, 20240627, 20240628, 20240629, 20240630, 20240631, 20240632, 20240633, 20240634, 20240635, 20240636, 20240637, 20240638, 20240639, 20240640, 20240641, 20240642, 20240643, 20240644, 20240645, 20240646, 20240647, 20240648, 20240649, 20240650, 20240651, 20240652, 20240653, 20240654, 20240655, 20240656, 20240657, 20240658, 20240659, 20240660, 20240661, 20240662, 20240663, 20240664, 20240665, 20240666, 20240667, 20240668, 20240669, 20240670, 20240671,…

    Quantity: 1661

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 96471
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.

More recalls from this firm