RecallRadar
FDA · Medical devicesClass I

Qapel Medical Inc. medical device recall: Hippo 072 Aspiration System including Cheetah Delivery Tool and…

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
96492
Recall initiated
Feb 26, 2025
Classified by FDA
Apr 7, 2025
Posted in Enforcement Report
Apr 16, 2025
Recalling firm
Qapel Medical Inc.
Firm location
Fremont, CA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Feb 26, 2025. FDA classified it 40 days later, on Apr 7, 2025. Distribution reported by FDA: US: CA, IL, NE, TN, TX, NJ, NC, SD, FL, MI, PA, GA, AL, IN, NY, WI, AZ, MN, LA, OH, OK, SC, NV, ND, ID, MO, CT, WA, MA. OUS: QATAR, United Arab Emirates, Republic of Kazakhstan

Recalled products (1)

  • Z-1512-2025 Class I Ongoing

    Hippo 072 Aspiration System including Cheetah Delivery Tool and Aspiration Tubing, REF: APT6072-132. Hippo 072 Aspiration System including Cheetah Delivery Tool, REF: AP6072-132. 072 Aspiration Tubing, REF: APT-95

    Codes / lots: REF/UDI-DI/Lot(Expiration): APT6072-132/00857545008127/FG241008C-03(04/07/2025), FG240916C-04(03/17/2025), FG240905C-04(03/06/2025). AP6072-132/00857545008097/FG241206A-03(06/08/2025), FG240917A-01(03/17/2025). APT-95/00857545008103/FG241206A-04(06/08/25)

    Quantity: 1617

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 96492
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.