RecallRadar
FDA · Medical devicesClass I

Covidien medical device recall: Newport HT70 Ventilator

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
96774
Recall initiated
May 21, 2025
Classified by FDA
Jun 9, 2025
Posted in Enforcement Report
Jun 18, 2025
Recalling firm
Covidien
Firm location
Carlsbad, CA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on May 21, 2025. FDA classified it 19 days later, on Jun 9, 2025. Distribution reported by FDA: Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece,… Covidien has 127 recalls in our records.

Recalled products (1)

  • Z-1915-2025 Class I Ongoing

    Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WW-NA. HT70-2 CONTROL BOARD ROHS, REF: GR-PCB3252A. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.

    Codes / lots: REF/UDI-DI/Serial Numbers: HT70M-JP-NA/00839935008221/N10HT700710132, N10HT700710135, N10HT700710136, N10HT700710159, N10HT700710161, N10HT700710162, N10HT700710163, N10HT700710164, N10HT700810194, N10HT700810199, N10HT700810202, N10HT700810204, N10HT700810205, N10HT700810206, N10HT700910223, N10HT700910227, N10HT701010279, N10HT701011015, N10HT701011016, N10HT701111044, N10HT701111048, N10HT701111049, N10HT701111050, N11HT700111137, N11HT700111139, N11HT700111141, N11HT700111142, N11HT700111143, N11HT700111144, N11HT700211145, N11HT700211155, N11HT700211160, N11HT700211161, N11HT700211163, N11HT700211164, N11HT700311168, N11HT700311169, N11HT700311170, N11HT700311183, N11HT700311184, N11HT700311185, N11HT700311186, N11HT700311188, N11HT700311189, N11HT700311191, N11HT700311192, N11HT700311193, N11HT700311195, N11HT700311196, N11HT700311197, N11HT700311201, N11HT700311202,…

    Quantity: 5394

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 96774
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.

More recalls from this firm