RecallRadar
FDA · Medical devicesClass I

Zyno Medical LLC medical device recall: Z-800 Infusion System. Intended to provide intravenous infusion of…

Software problems

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
96796
Recall initiated
May 7, 2025
Classified by FDA
Jun 6, 2025
Posted in Enforcement Report
Jun 18, 2025
Recalling firm
Zyno Medical LLC
Firm location
Natick, MA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
E-Mail
Products in this event
4

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Unreleased software versions were installed on distributed devices without verification or validation.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on May 7, 2025. FDA classified it 30 days later, on Jun 6, 2025. This recall event covers 4 separate products, each with its own FDA recall number. Distribution reported by FDA: US Nationwide Zyno Medical LLC has 9 recalls in our records.

Recalled products (4)

  • Z-1865-2025 Class I Ongoing

    Z-800 Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

    Codes / lots: Lot Code: Model Number; Z800; UDI-DI: 00814377102006; Serial No.: 800242; 801925; 802867; 900875; 901058; 901089; 901101; 901126; 901143; 901567; 904021; 904073; 800378; 801950; 803020; 900914; 901061; 901090; 901102; 901129; 901145; 901795; 904023; 904083; 801082; 802113; 803268; 900961; 901066; 901091; 901103; 901130; 901230; 904004; 904026; 904087; 801353; 802131; 803526; 900964; 901070; 901092; 901110; 901131; 901240; 904006; 904027; 904093; 801420; 802160; 803768; 901001; 901076; 901093; 901112; 901132; 901241; 904007; 904028; 904094; 801503; 802645; 900029; 901014; 901077; 901094; 901113; 901133; 901244; 904009; 904029; 905077; 801601; 802753; 900035; 901024; 901081; 901095; 901115; 901134; 901245; 904011; 904030; 905162; 801640; 802764; 900051; 901041; 901083; 901096; 901116; 901136; 901246; 904012; 904031; 906681; 801852; 802771; 900115; 901054; 901085; 901097; 901117; 901138;…

    Quantity: 613 units

  • Z-1866-2025 Class I Ongoing

    Z-800F Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

    Codes / lots: Model No: Z800F; UDI-DI: 00814371020013; Serial No. 600091; 600133; 600203; 600259; 600320; 600361; 600404; 600446; 600482; 600526; 600565; 600601; 600094; 600138; 600204; 600261; 600321; 600362; 600406; 600447; 600483; 600528; 600566; 600603; 600095; 600139; 600206; 600262; 600322; 600363; 600407; 600449; 600485; 600529; 600568; 600604; 600099; 600141; 600207; 600263; 600323; 600364; 600408; 600450; 600487; 600531; 600569; 600605; 600100; 600142; 600209; 600265; 600324; 600369; 600410; 600451; 600488; 600532; 600570; 600608; 600102; 600143; 600211; 600266; 600325; 600370; 600411; 600452; 600489; 600533; 600571; 600609; 600104; 600146; 600215; 600267; 600326; 600376; 600412; 600453; 600490; 600535; 600572; 600610; 600105; 600148; 600217; 600268; 600328; 600377; 600414; 600454; 600495 600536; 600573; 600611; 600106; 600150; 600220; 600270; 600330; 600378; 600415; 600456; 600496; 600539;…

    Quantity: 613 units

  • Z-1867-2025 Class I Ongoing

    Z-800W Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

    Codes / lots: Model Number: Z800W; UDI-DI; 00814371020020; Serial No. 700063; 700163; 700203; 700402; 700488; 700093; 700169; 700210; 700428; 700495; 700104; 700172; 700214; 700431; 700503; 700111; 700188; 700225; 700440; 700510; 700120; 700189; 700230; 700449; 700513; 700135; 700190; 700285; 700460; 700541; 700145; 700192; 700375; 700471; 700481; 700150; 700195; 700391; 700472; 700397; 700159; 700199; 700393; 700478; 700201; 700162.

    Quantity: 613 units

  • Z-1868-2025 Class I Ongoing

    Z-800WF Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

    Codes / lots: Model No. Z800WF; UDI-DI: 00814371020037; Serial No. 500004; 500005; 500006; 500810; 500812; 500813; 501748; 503680; 503681; 503682.

    Quantity: 613 units

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 96796
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.