mo-Vis BVBA medical device recall: Micro Joystick R-net. Electrical wheelchair component
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 97354
- Recall initiated
- Jul 10, 2025
- Classified by FDA
- Sep 2, 2025
- Posted in Enforcement Report
- Sep 10, 2025
- Recalling firm
- mo-Vis BVBA
- Firm location
- Nevele, Belgium
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Products in this event
- 6
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Jul 10, 2025. FDA classified it 54 days later, on Sep 2, 2025. This recall event covers 6 separate products, each with its own FDA recall number. Distribution reported by FDA: US Nationwide distribution in the state of TX. mo-Vis BVBA has 2 recalls in our records.
Recalled products (6)
Z-2454-2025 Class I Ongoing
Micro Joystick R-net. Electrical wheelchair component.
Codes / lots: Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.
Quantity: 3 units
Z-2455-2025 Class I Ongoing
Multi Joystick R-net. Electrical wheelchair component.
Codes / lots: Model No. P002-62; UDI-DI: 05407008320041; Serial No. 1000 to 1690.
Quantity: 3 units
Z-2456-2025 Class I Ongoing
IDM-MICRO-R. Electrical wheelchair component.
Codes / lots: Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.
Quantity: 15 units
Z-2457-2025 Class I Ongoing
IDM-MULTI-R. Electrical wheelchair component.
Codes / lots: Model No. PRSPS0016; UDI-DI: 05407008320850; Serial No. 1000 to 1690.
Quantity: 15 units
Z-2458-2025 Class I Ongoing
All-round Joystick R-net Light. Electrical wheelchair component.
Codes / lots: Model No. P002-76; UDI-DI: 05407008320096; Serial No. 1000 to 2647.
Quantity: 3 units
Z-2459-2025 Class I Ongoing
IDM-ARLITE-R. Electrical wheelchair component.
Codes / lots: Model No. PRSPS00018. UDI-DI: 05407008320102; Serial No. 1000 to 2647.
Quantity: 25 units
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
- Official notice
- FDA Enforcement Report, event 97354
- Raw record
- openFDA API (JSON)
- Dataset last updated by FDA
- Nov 3, 2026
- Imported by RecallRadar
- Nov 4, 2026
Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.