RecallRadar
FDA · Medical devicesClass I

CareFusion 303, Inc. medical device recall: BD Alaris Pump Infusion Sets for the following reference numbers:…

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
97519
Recall initiated
Sep 11, 2025
Classified by FDA
Oct 8, 2025
Posted in Enforcement Report
Oct 15, 2025
Recalling firm
CareFusion 303, Inc.
Firm location
San Diego, CA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Sep 11, 2025. FDA classified it 27 days later, on Oct 8, 2025. Distribution reported by FDA: US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana Canada Switzerland China France United Kingdom Gibraltar Israel India Kuwait Mexico Malaysia Netherlands New Zealand Philippines Pakistan Qatar Saudi Arabia Singapore Taiwan, Province of China South Africa CareFusion 303, Inc. has 99 recalls in our records.

Recalled products (1)

  • Z-0035-2026 Class I Ongoing

    BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve SmartSite Port (Burette) 2 SmartSites Y-Sites 24201-0007, 2 SmartSite Y-sites 10012645, Back Check Valve, Non-vented, SmartSite Y-site 2426-0500, Back Check Valve, 3 SmartSite Y-sites 2450-0500, Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSite Y-sites 11522558, Ball Valve, Back Check Valve, 3 SmartSite Y-sites 2410-0500, 2 SmartSite Y-Sites 2232-0007, 0.2 Micron Filter 11613191, 0.2 Micron Filter, Ball Valve, SmartSite" Port (Burette), 2 SmartSite Y-sites 10012144, 3 SmartSite Y-Sites, Check Valve 10013034, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 4-way Stopcock 10013037, Non-Vented, 180 Micron Filter, Low Sorbing Tubing Segment, 1 SmartSite, Y-Site 11582773, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 3-way Stopcock

    Codes / lots: All Lots/ UDI: C24101E - 10885403235528, 10015414 - 10885403233999, 2447-0007 - 10885403235078, 24201-0007 - 10885403476341, 10012645 - 10885403233753, 2426-0500 - 7613203020992, 11522558 - 10885403232305, 2410-0500 - 7613203012492, 2232-0007 - 10885403463969, 11613191 - 10885403276026, 10012144 - 7613203021159, 10013034 - 10885403232312, 10013037 - 10885403233784, 11582773 - 10885403239588

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 97519
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.