RecallRadar
FDA · Medical devicesClass I

Balt USA, LLC medical device recall: Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega…

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
97661
Recall initiated
Sep 25, 2025
Classified by FDA
Oct 24, 2025
Posted in Enforcement Report
Nov 5, 2025
Recalling firm
BALT USA, LLC
Firm location
Irvine, CA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Sep 25, 2025. FDA classified it 29 days later, on Oct 24, 2025. Distribution reported by FDA: US distribution to states of: AL, CA, CO, CT, FL, IL, KY, MA, NY, TN, TX, and UT. BALT USA, LLC has 6 recalls in our records.

Recalled products (1)

  • Z-0188-2026 Class I Ongoing

    Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast incorporates a Luer fitting hub for the attachment of accessories. The Mega Ballast is provided sterile, non-pyrogenic, and is intended for single use only. Component: Not a component

    Codes / lots: Model Numbers and UDI-DI codes: Model No MEGABT09180 (00840303711780) Model No MEGABT09190 (00840303711797) Model No MEGABT091100 (00840303711803) Model No MEGABT091110 (00840303711810) Lot Numbers: F250600687, F250600686, F250600685, F250600455, F250600346, F250600345, F250600010, F250501434, F250501346, F250501091, F250500824, F250400219, F250400024, F250301020, F250300297, F250100956

    Quantity: 207 catheters

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 97661
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.