RecallRadar
FDA · Medical devicesClass I

IMPRIS Imaging Inc medical device recall: IMRIS Operating Suite

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
97750
Recall initiated
Sep 30, 2025
Classified by FDA
Oct 28, 2025
Posted in Enforcement Report
Nov 5, 2025
Recalling firm
IMPRIS Imaging Inc
Firm location
Chaska, MN, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
E-Mail
Products in this event
4

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Sep 30, 2025. FDA classified it 28 days later, on Oct 28, 2025. This recall event covers 4 separate products, each with its own FDA recall number. Distribution reported by FDA: Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.

Recalled products (4)

  • Z-0161-2026 Class I Ongoing

    IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio

    Codes / lots: IMRIS Part Number(s): 105118-000, UDI/DI None predates UDI; 105118-001, UDI/DI None predates UDI; 105117-003, UDI/DI None predates UDI; 105117-004, UDI/DI None predates UDI; 105048-000, UDI/DI None predates UDI; 105048-001, UDI/DI None predates UDI; 105103-001, UDI/DI None predates UDI; 105071-004, UDI/DI None predates UDI; 105824-000, UDI/DI 00857534006189; 105828-000, UDI/DI 00857534006165; 116189-000, UDI/DI 00857534006202; 118396-000, UDI/DI 00857534006196.

  • Z-0162-2026 Class I Ongoing

    IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra

    Codes / lots: IMRIS Part Number(s): 116053-000, UDI/DI 00857534006035; 116117-000, UDI/DI 00857534006042; 116118-000, UDI/DI 00857534006059; 116119-000, UDI/DI 00857534006066.

  • Z-0163-2026 Class I Ongoing

    IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA

    Codes / lots: IMRIS Part Number(s): 120823-000, UDI/DI 00857534006745

  • Z-0164-2026 Class I Ongoing

    IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida

    Codes / lots: IMRIS Part Number(s): 120823-000, UDI/DI 00857534006806

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 97750
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.