RecallRadar
FDA · Medical devicesClass I

Siemens Medical Solutions USA, Inc medical device recall: MAGNETOM Cima.X Upgrade. Model Number: 11689304

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
97845
Recall initiated
Oct 1, 2025
Classified by FDA
Nov 12, 2025
Posted in Enforcement Report
Nov 19, 2025
Recalling firm
Siemens Medical Solutions USA, Inc
Firm location
Malvern, PA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Oct 1, 2025. FDA classified it 42 days later, on Nov 12, 2025. Distribution reported by FDA: US distribution to California. International distribution to Australia, Germany, Great Britian. Siemens Medical Solutions USA, Inc has 400 recalls in our records.

Recalled products (1)

  • Z-0431-2026 Class I Ongoing

    MAGNETOM Cima.X Upgrade. Model Number: 11689304.

    Codes / lots: Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.

    Quantity: 4 system (1 US; 3 OUS)

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 97845
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.