RecallRadar
FDA · Medical devicesClass I

Baxter Healthcare Corporation medical device recall: Hillrom VOLARA System P.CIRCUIT KIT

Lead

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
99119
Recall initiated
May 21, 2026
Classified by FDA
Jul 1, 2026
Posted in Enforcement Report
Jul 8, 2026
Recalling firm
Baxter Healthcare Corporation
Firm location
Deerfield, IL, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
4

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on May 21, 2026. FDA classified it 41 days later, on Jul 1, 2026. This recall event covers 4 separate products, each with its own FDA recall number. Distribution reported by FDA: Worldwide distribution - US Nationwide and the country of Canada. Baxter Healthcare Corporation has 279 recalls in our records.

Recalled products (4)

  • Z-2548-2026 Class I Ongoing

    Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

    Codes / lots: UDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/2025

    Quantity: 10540 units

  • Z-2549-2026 Class I Ongoing

    Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device

    Codes / lots: UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025

    Quantity: 0

  • Z-2550-2026 Class I Ongoing

    Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device

    Codes / lots: UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025

    Quantity: 16187 units

  • Z-2551-2026 Class I Ongoing

    Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device

    Codes / lots: UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/2025

    Quantity: 13881 units

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 99119
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.

More recalls from this firm