Nova Biomedical Corporation medical device recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose…
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 99546
- Recall initiated
- Jul 29, 2026
- Classified by FDA
- Sep 11, 2026
- Posted in Enforcement Report
- Sep 23, 2026
- Recalling firm
- Nova Biomedical Corporation
- Firm location
- Waltham, MA, United States
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Products in this event
- 5
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Jul 29, 2026. FDA classified it 44 days later, on Sep 11, 2026. This recall event covers 5 separate products, each with its own FDA recall number. Distribution reported by FDA: Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom. Nova Biomedical Corporation has 7 recalls in our records.
Recalled products (5)
Z-3128-2026 Class I Ongoing
StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.
Codes / lots: Model No: 63685. UDI-DI: 00385480636858. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Quantity: 20,616 units
Z-3129-2026 Class I Ongoing
StatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.
Codes / lots: Model No: 66739. UDI-DI: 0038548063685. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Quantity: 10,941 units
Z-3130-2026 Class I Ongoing
StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.
Codes / lots: Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Quantity: 1,858 units
Z-3131-2026 Class I Ongoing
StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.
Codes / lots: Model No: 63910. UDI-DI: 00385480639101. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Quantity: 8,939 units
Z-3132-2026 Class I Ongoing
StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.
Codes / lots: Model No: 66429. UDI-DI: :00385480664295. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Quantity: 2,044 units
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
- Official notice
- FDA Enforcement Report, event 99546
- Raw record
- openFDA API (JSON)
- Dataset last updated by FDA
- Nov 3, 2026
- Imported by RecallRadar
- Nov 4, 2026
Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.