Nova Biomedical Corporation medical device recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose…

Software problems

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
99546
Recall initiated
Jul 29, 2026
Classified by FDA
Sep 11, 2026
Posted in Enforcement Report
Sep 23, 2026
Recalling firm
Nova Biomedical Corporation
Firm location
Waltham, MA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
E-Mail
Products in this event
5

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Jul 29, 2026. FDA classified it 44 days later, on Sep 11, 2026. This recall event covers 5 separate products, each with its own FDA recall number. Distribution reported by FDA: Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom. Nova Biomedical Corporation has 7 recalls in our records.

Recalled products (5)

  • Z-3128-2026 Class I Ongoing

    StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.

    Codes / lots: Model No: 63685. UDI-DI: 00385480636858. Software Versions: All versions 0.00.13.10 through 1.5.14.25.

    Quantity: 20,616 units

  • Z-3129-2026 Class I Ongoing

    StatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.

    Codes / lots: Model No: 66739. UDI-DI: 0038548063685. Software Versions: All versions 0.00.13.10 through 1.5.14.25.

    Quantity: 10,941 units

  • Z-3130-2026 Class I Ongoing

    StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.

    Codes / lots: Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25.

    Quantity: 1,858 units

  • Z-3131-2026 Class I Ongoing

    StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.

    Codes / lots: Model No: 63910. UDI-DI: 00385480639101. Software Versions: All versions 0.00.13.10 through 1.5.14.25.

    Quantity: 8,939 units

  • Z-3132-2026 Class I Ongoing

    StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.

    Codes / lots: Model No: 66429. UDI-DI: :00385480664295. Software Versions: All versions 0.00.13.10 through 1.5.14.25.

    Quantity: 2,044 units

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 99546
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.

More recalls from this firm