CooperSurgical, Inc. medical device recall: Inca
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 99643
- Recall initiated
- Aug 11, 2026
- Classified by FDA
- Sep 18, 2026
- Posted in Enforcement Report
- Sep 30, 2026
- Recalling firm
- CooperSurgical, Inc.
- Firm location
- Trumbull, CT, United States
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Letter
- Products in this event
- 2
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Aug 11, 2026. FDA classified it 38 days later, on Sep 18, 2026. This recall event covers 2 separate products, each with its own FDA recall number. Distribution reported by FDA: Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru. CooperSurgical, Inc. has 58 recalls in our records.
Recalled products (2)
Z-3139-2026 Class I Ongoing
INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.
Codes / lots: 1. Model Number: 44-2707. UDI-DI: (01)20888937009433. Lot Code: 550019722. 2. Model Number: 44-2709. UDI-DI: (01)60888937009448. Lot Codes: 550031617, 550033781, 550031621. 3. Model Number: 44-2710. UDI-DI: (01)20888937009457. Lot Codes: 550031615, 550031616, 550031620, 550034760. 4. Model Number: 44-2712. UDI-DI: (01)20888937009464. Lot Codes: 550031611, 550031614, 550033779, 550031619. 5. Model Number: 44-2715. UDI-DI: (01)20888937009471. Lot Codes: 550031618, 550033780, 550034763.
Quantity: 2,643 boxes (13,215 units)
Z-3140-2026 Class I Ongoing
INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.
Codes / lots: 1. Model Number: 44-0712. UDI-DI: (01)60888937009516. Lot Code: 550031624. 2. Model Number: 44-0707. UDI-DI: (01)20888937009488. Lot Code: 550021057. 3. Model Number: 44-0709. UDI-DI: (01)20888937009495. Lot Code: 550031623. 4. Model Number: 44-0715. UDI-DI: (01)20888937009525. Lot Code: 550031625.
Quantity: 149 boxes (745 units)
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
- Official notice
- FDA Enforcement Report, event 99643
- Raw record
- openFDA API (JSON)
- Dataset last updated by FDA
- Nov 3, 2026
- Imported by RecallRadar
- Nov 4, 2026
Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.