Medline Industries, LP medical device recall: Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 99827
- Recall initiated
- Aug 14, 2026
- Classified by FDA
- Sep 30, 2026
- Posted in Enforcement Report
- Oct 7, 2026
- Recalling firm
- Medline Industries, LP
- Firm location
- Northfield, IL, United States
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Letter
- Products in this event
- 2
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Aug 14, 2026. FDA classified it 47 days later, on Sep 30, 2026. This recall event covers 2 separate products, each with its own FDA recall number. Distribution reported by FDA: Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama. Medline Industries, LP has 257 recalls in our records.
Recalled products (2)
Z-3227-2026 Class I Ongoing
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
Codes / lots: MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots
Quantity: 771,127 total
Z-3228-2026 Class I Ongoing
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE
Codes / lots: MDR107000P UDI-DI 00080196288284 All lots
Quantity: 771,127 total
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
- Official notice
- FDA Enforcement Report, event 99827
- Raw record
- openFDA API (JSON)
- Dataset last updated by FDA
- Nov 3, 2026
- Imported by RecallRadar
- Nov 4, 2026
Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.