Hamilton Medical AG recalls
12 recalls in our records from 2018 to 2026: 12 FDA medical devices.
Recalls with full detail pages
-
Hamilton Medical AG medical device recall: Breathing circuit set
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result…
-
Hamilton Medical AG medical device recall: Hamilton C6 REF 160021
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient…
-
Hamilton Medical AG medical device recall: Breathing circuit set
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and…
Earlier and other recalls
These link to the official agency notice.
-
Hamilton Medical AG medical device recall: IntelliCuff Official notice ↗
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff…
-
Hamilton Medical AG medical device recall: Ventilator HAMILTON-C6 Official notice ↗
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2…
-
Hamilton Medical AG medical device recall: Hamilton-C6 Official notice ↗
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored,…
-
Hamilton Medical AG medical device recall: Hamilton-C6 Official notice ↗
Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight…
-
Hamilton Medical AG medical device recall: HAMILTON-C6 Intensive Care Ventilator Official notice ↗
Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults…
-
Hamilton Medical AG medical device recall: HAMILTON-H900 Humidifier Official notice ↗
When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not…
-
Hamilton Medical AG medical device recall: Hamilton-G5 Official notice ↗
New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves…
-
Hamilton Medical AG medical device recall: Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator… Official notice ↗
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down…
-
Hamilton Medical AG medical device recall: Hamilton-C2 Ventilator Unit Intended to provide ventilator support to… Official notice ↗
After two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
Firm names are matched across agencies by name only. Different companies with similar names may be grouped together; check each official notice.