RecallRadar
FDA · Medical devicesClass I

Hamilton Medical AG medical device recall: Hamilton C6 REF 160021

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
97079
Recall initiated
Jun 3, 2025
Classified by FDA
Aug 19, 2025
Posted in Enforcement Report
Aug 27, 2025
Recalling firm
Hamilton Medical AG
Firm location
Domat/Ems, Switzerland
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Letter
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Jun 3, 2025. FDA classified it 77 days later, on Aug 19, 2025. Distribution reported by FDA: US Nationwide distribution. Hamilton Medical AG has 12 recalls in our records.

Recalled products (1)

  • Z-2328-2025 Class I Ongoing

    Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

    Codes / lots: Model/Catalog Number: 160021 ESM board Part Number: 10144626 GTIN: 07630002808590 UDI-DI: 76300028PN160021ZW Affected SN range: 13636 till 13950 and 13966 till 14753 13797 13801 13806 13855 13865 13981 13988 13989 14012 14015 14019 14034 14035 14042 14052 14054 14102 14109 14112 14113 14114 14115 14116 14119 14120 14121 14122 14123 14124 14126 14127 14128 14163 14168 14169 14170 14172 14173 14175 14177 14187 14188 14189 14195 14197

    Quantity: 1040 systems

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 97079
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.