Fresenius Kabi USA, LLC medical device recall: Ivenix Infusion System
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 95891
- Recall initiated
- Dec 6, 2024
- Classified by FDA
- Jan 8, 2025
- Posted in Enforcement Report
- Jan 15, 2025
- Recalling firm
- Fresenius Kabi USA, LLC
- Firm location
- North Andover, MA, United States
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Products in this event
- 1
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Dec 6, 2024. FDA classified it 33 days later, on Jan 8, 2025. Distribution reported by FDA: US distribution to states of: ID, MN, NV, TX, VA, WI. Fresenius Kabi USA, LLC has 67 recalls in our records.
Recalled products (1)
Z-0763-2025 Class I Ongoing
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Codes / lots: Model Number: LVP-0004, UDI Number: 00811505030320. Serial Numbers: 9233532106, 9233421830, 9233421828, 9233532158, 9233522076, 9233522041, 9233421833, 9233421829, 9233421808, 9233421654, 9233421626, 9233421621, 9233321340, 2215401517, 2215401545, 9233421789, 9233421809, 9233532167, 9233532100, 9233522070, 9233421825, 9233421824, 9233421794, 9233421783, 9233421781, 9233421771, 9233421768, 9233421756, 9233421736, 9233421713, 9233421702, 9233421617, 9233421611, 9233391513, 9233391491, 9233532098, 9233421822, 9233421797, 9233421678, 9233421666, 9233421601, 9233532114, 9233421792, 9233421766, 9233421687, 2215401372, 9233522040, 9233522037, 9233421764, 9233421752, 9233421665, 9232490602, 9233522062, 9233421817, 9233421832, 9233522094, 9233421805, 9233421740, 9233532161, 9233532111, 9233522061, 9233421836, 2212200949, 2212200956, 2212200959, 2212200960, 2212200964, 2212200965, 2212200983,…
Quantity: 1,520 units
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset