RecallRadar
FDA · Medical devicesClass I

Fresenius Kabi USA, LLC medical device recall: Ivenix Infusion System

Software problems

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
96085
Recall initiated
Jan 10, 2025
Classified by FDA
Feb 4, 2025
Posted in Enforcement Report
Feb 12, 2025
Recalling firm
Fresenius Kabi USA, LLC
Firm location
North Andover, MA, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Products in this event
1

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Jan 10, 2025. FDA classified it 25 days later, on Feb 4, 2025. Distribution reported by FDA: US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI. Fresenius Kabi USA, LLC has 67 recalls in our records.

Recalled products (1)

  • Z-1019-2025 Class I Ongoing

    Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

    Codes / lots: UDI-DI: 00811505030122. Software 5.9.2 and prior

    Quantity: 23 systems

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 96085
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.