Baxter Healthcare Corporation medical device recall: Baxter Novum IQ Large Volume Pump
Key facts
- Agency
- U.S. Food and Drug Administration (FDA)
- Program
- Medical devices
- Classification
- Class I
- Status
- Ongoing
- Event ID
- 97164
- Recall initiated
- Jul 14, 2025
- Classified by FDA
- Aug 14, 2025
- Posted in Enforcement Report
- Aug 20, 2025
- Recalling firm
- Baxter Healthcare Corporation
- Firm location
- Deerfield, IL, United States
- Initiated by
- Voluntary: Firm initiated
- Firm notified customers by
- Letter
- Products in this event
- 1
What to do
Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.
Reason for recall
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.
What the classification means
- Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.
About this recall
The firm started this recall on Jul 14, 2025. FDA classified it 31 days later, on Aug 14, 2025. Distribution reported by FDA: Worldwide distribution - US Nationwide and the country of Canada. Baxter Healthcare Corporation has 279 recalls in our records.
Recalled products (1)
Z-2173-2025 Class I Ongoing
Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS
Codes / lots: UDI/DI 05413765851797, All Serial Numbers
Quantity: 36705 units
Source and freshness
- Source
- U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
- Official notice
- FDA Enforcement Report, event 97164
- Raw record
- openFDA API (JSON)
- Dataset last updated by FDA
- Nov 3, 2026
- Imported by RecallRadar
- Nov 4, 2026
Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.