RecallRadar
FDA · Medical devicesClass I

Medline Industries, LP medical device recall: Anesthesia Circuit Kits and other various kits that contain…

Lead

Key facts

Agency
U.S. Food and Drug Administration (FDA)
Program
Medical devices
Classification
Class I
Status
Ongoing
Event ID
97882
Recall initiated
Oct 10, 2025
Classified by FDA
Nov 13, 2025
Posted in Enforcement Report
Nov 19, 2025
Recalling firm
Medline Industries, LP
Firm location
Northfield, IL, United States
Initiated by
Voluntary: Firm initiated
Firm notified customers by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Products in this event
10

What to do

Compare the product, lot or serial numbers below with what you have. If you use a prescription drug or medical device in this recall, talk to your pharmacist or health care provider before you stop using it, and follow the firm's instructions.

View the official FDA record ↗

Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

What the classification means

  • Class I: FDA's most serious category: there is a reasonable probability that using or being exposed to the product will cause serious health problems or death.

About this recall

The firm started this recall on Oct 10, 2025. FDA classified it 34 days later, on Nov 13, 2025. This recall event covers 10 separate products, each with its own FDA recall number. Distribution reported by FDA: International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI; Medline Industries, LP has 257 recalls in our records.

Recalled products (10)

  • Z-0442-2026 Class I Ongoing

    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;

    Codes / lots: DYNJAA0309A UDI-DI 40193489400268 (case), 10193489400267 (ea) Lots 25EBU353 25GBI473; DYNJAA0360 UDI DI 40193489419000 (case), 10193489419009 (ea) Lots 25EBU304 25GBQ732; DYNJAA0368A UDI-DI 40195327092550 (case), 10195327092559 (ea) Lots 25GBH582; DYNJAA0380A UDI-DI 40195327478286 (case), 10195327478285 (ea) Lots 25FBB950 25FBT875 25GBX623; DYNJAA0433 UDI-DI 40198459320706 (case), 10198459320705 (ea) Lots 25EBH621 25EBQ391; DYNJAA10261B UDI- DI 40193489430784 (case), 10193489430783 (ea) Lots 25FBJ760; DYNJAA10368A UDI-DI 40198459320720 (case), 10198459320729 (ea) Lots 25GBY374 25IBK072; DYNJAA10392 UDI-DI 40888277962201 (case), 10888277962200 (ea) Lots 25FBN032; DYNJAA10430A UDI-DI 40193489832052 (case), 10193489832051 (ea) Lots 25FBW335 25IBH048; DYNJAA10547A UDI-DI 40193489603188 (case), 10193489603187 (ea) 25FBT853; DYNJAA10591 UDI-DI 40889942632535 (case), 10889942632534 (ea) Lots…

    Quantity: 96,406 total units

  • Z-0443-2026 Class I Ongoing

    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550

    Codes / lots: DYNJ10550 UDI-DI 40889942469704 (case), 10889942469703 (ea) Lot 25EBQ445

    Quantity: 96,406 total units

  • Z-0444-2026 Class I Ongoing

    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F

    Codes / lots: DYNJ909116D UDI-DI 40198459055356 (case), 10198459055355 (ea) Lots 25FBM604 25GBP604; DYNJ909124F UDI-DI 40198459130619 (case), 10198459130618 (ea) Lots 25FBB831 25GBI833

    Quantity: 96,406 total units

  • Z-0445-2026 Class I Ongoing

    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D

    Codes / lots: ENT DYNJ909128I UDI-DI 40198459333355 (case), 10198459333354 (ea) Lots 25GBO685; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D UDI-DI 40198459055301 (case), 10198459055300 (ea) Lots 25GBN602

    Quantity: 96,406 total units

  • Z-0446-2026 Class I Ongoing

    DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF

    Codes / lots: CYSTO DYNJ909127F UDI-DI 40198459055424 (case), 10198459055423 (ea) Lots 25FBP819 25GBM605; GENERAL LAPAROSCOPY-LF DYNJ909129D UDI-DI 40198459055264 (case), 10198459055263 (ea) Lots 25FBV537; ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909119G UDI-DI 40198459130527 (case), 10198459130526 (ea) Lots 25FBP681

    Quantity: 96,406 total units

  • Z-0447-2026 Class I Ongoing

    Convenience Kit, BREAST HERNIA DYNJ909125G

    Codes / lots: DYNJ909125G UDI-DI 40198459130664 (case), 10198459130663 (ea) Lot 25FBU543 25IBD459

    Quantity: 96,406 total units

  • Z-0448-2026 Class I Ongoing

    Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING

    Codes / lots: HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486; ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070; SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708; UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707

    Quantity: 96,406 total units

  • Z-0449-2026 Class I Ongoing

    Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230

    Codes / lots: CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230 UDI-DI 40884389688589 (case), 10884389688588 (ea) Lots 25EBV570 25FBT843

    Quantity: 96,406 total units

  • Z-0450-2026 Class I Ongoing

    Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE

    Codes / lots: GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYNJ909120D UDI-DI 40198459130541 (case), 10198459130540 (ea) Lots 25GBT821 25IBF809

    Quantity: 96,406 total units

  • Z-0451-2026 Class I Ongoing

    Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY

    Codes / lots: DR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788

    Quantity: 96,406 total units

Source and freshness

Source
U.S. Food and Drug Administration (FDA) Enforcement Report, via the openFDA device enforcement dataset
Official notice
FDA Enforcement Report, event 97882
Raw record
openFDA API (JSON)
Dataset last updated by FDA
Nov 3, 2026
Imported by RecallRadar
Nov 4, 2026

Enforcement Report entries describe recalls FDA has classified. They are not always the same as a company press release. If something here differs from FDA's record, FDA's record is correct — please tell us.